Goren, AC; Bilsel, G; Bilsel, M; Yenisoy-Karakas, S; Karakas, D
{ "DOI": "10.1016/j.chroma.2004.09.048", "abstract": "Atropine and obidoxime in a parenteral injection device are determined by simple HPLC method simultaneously without any pretreatment at 228 nm. The relative standard deviations (R.S.D.) were below 1.6% for the compounds. The correlation coefficient was greater than 0.999 for both compounds in the calibration range. The recoveries at 5 mg/L concentration averaged as 95% for atropine and 102% for obidoxime. The uncertainty of the measurements for atropine and obidoxime was 2.8% and 2.4%, respectively. (C) 2004 Elsevier B.V. All rights reserved.", "author": [ { "family": "Goren", "given": " AC" }, { "family": "Bilsel", "given": " G" }, { "family": "Bilsel", "given": " M" }, { "family": "Yenisoy-Karakas", "given": " S" }, { "family": "Karakas", "given": " D" } ], "container_title": "JOURNAL OF CHROMATOGRAPHY A", "id": "94623", "issue": "1-2", "issued": { "date-parts": [ [ 2004, 1, 1 ] ] }, "page": "237-239", "title": "Simple high-performance liquid chromatographic method for determination of atropine and obidoxime in a parenteral injection device", "type": "article-journal", "volume": "1057" }