Dergi makalesi Açık Erişim
Armutcu, Canan; Uzun, Lokman; Denizli, Adil
Automated on-line two-dimensional high-performance liquid chromatography method (2D-HPLC) is proposed to determine Ochratoxin A (OTA) in food samples as an alternative to OTA immunoaffinity column (IAC). An on-line 2D-HPLC system is designed for the analysis of OTA using an affinity-based monolithic column in the first dimension and reversed-phase C18 column in the second dimension. Initially, optimal OTA separation efficiency is determined through traditional HPLC system consisting of a P(HEMAPA) monolithic column coupled with HPLC system. Secondly, after providing optimum conditions, OTA determination was investigated through the 2D-HPLC system. According to results, 2D-HPLC system showed good linearity in the range 0.5 to 20 ng/mL with limit of detection (LOD) and limit of quantification (LOQ) values of 21.2 pg/mL and 64.3 pg/mL, respectively. The P(HEMAPA)-4 monolithic column displayed good recovery of OTA ranging from 104.34% to 107.33%. Relative standard deviations (RSD) varied in the range 0.21% to 1.31% thus indicating the efficiency of P(HEMAPA)-4 monolithic column developed for OTA. (C) 2018 Elsevier B.V. All rights reserved.
Dosya adı | Boyutu | |
---|---|---|
bib-8f146e7a-3e3b-47d6-aea0-38a2205cb706.txt
md5:1e9417b9ce2fce0f34ba44cc64c6591d |
310 Bytes | İndir |
Görüntülenme | 13 |
İndirme | 3 |
Veri hacmi | 930 Bytes |
Tekil görüntülenme | 13 |
Tekil indirme | 3 |