Dergi makalesi Açık Erişim

Realization of luminous flux unit lumen at National Metrology Institute of Turkey (UME)

Samedov, F; Durak, M


Citation Style Language JSON

{
  "URL": "https://aperta.ulakbim.gov.tr/record/93741", 
  "abstract": "Luminous flux calibration facility with an integrating sphere has been installed at the National Metrology Institute of Turkey (UME) in order to realize lumen. A LabView-based computer -controlled electrical system has been developed to operate light sources and to control electrical parameters. Owing to the corrections of the sphere system, inner coating reflectance, spatial non-uniformity, self-absorption factor and interior temperature variations have been characterized. The overall uncertainty in luminous flux measurement has been determined, with an uncertainty of 1.14% (k = 2) in the range between 5-5000 lm.", 
  "author": [
    {
      "family": "Samedov", 
      "given": " F"
    }, 
    {
      "family": "Durak", 
      "given": " M"
    }
  ], 
  "container_title": "OPTICA APPLICATA", 
  "id": "93741", 
  "issue": "2", 
  "issued": {
    "date-parts": [
      [
        2004, 
        1, 
        1
      ]
    ]
  }, 
  "page": "265-274", 
  "title": "Realization of luminous flux unit lumen at National Metrology Institute of Turkey (UME)", 
  "type": "article-journal", 
  "volume": "34"
}
133
10
görüntülenme
indirilme
Görüntülenme 133
İndirme 10
Veri hacmi 1.5 kB
Tekil görüntülenme 131
Tekil indirme 10

Alıntı yap

Samedov, F. ve Durak, M. (2004). Realization of luminous flux unit lumen at National Metrology Institute of Turkey (UME). OPTICA APPLICATA, 34(2), 265–274. https://aperta.ulakbim.gov.tr/record/93741 adresinden erişildi.