Konferans bildirisi Açık Erişim

Construction of Maxwell-Wien Bridge at SASO-NMCC

Alshamrani, O. A.; Alrobaish, A. M.; Alqahtani, M. M.; Altakrouni, M. S.; Turhan, E.; Gulmez, G.


Dublin Core

<?xml version='1.0' encoding='utf-8'?>
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
  <dc:creator>Alshamrani, O. A.</dc:creator>
  <dc:creator>Alrobaish, A. M.</dc:creator>
  <dc:creator>Alqahtani, M. M.</dc:creator>
  <dc:creator>Altakrouni, M. S.</dc:creator>
  <dc:creator>Turhan, E.</dc:creator>
  <dc:creator>Gulmez, G.</dc:creator>
  <dc:date>2018-01-01</dc:date>
  <dc:description>An inductance calibration system for standard inductors based on a Maxwell-Wien bridge has been constructed at Saudi Standards Metrology and Quality Organization, National Measurement and Calibration Center (SASO NMCC). Expanded measurement uncertainty has been estimated to be less than 20 mu H/H for 100 mH inductance at 1 kHz. A bilateral comparison between SASO NMCC and TUBITAK UME has been performed in order to evaluate the performance of the calibration system.</dc:description>
  <dc:identifier>https://aperta.ulakbim.gov.trrecord/93097</dc:identifier>
  <dc:identifier>oai:zenodo.org:93097</dc:identifier>
  <dc:rights>info:eu-repo/semantics/openAccess</dc:rights>
  <dc:rights>http://www.opendefinition.org/licenses/cc-by</dc:rights>
  <dc:title>Construction of Maxwell-Wien Bridge at SASO-NMCC</dc:title>
  <dc:type>info:eu-repo/semantics/conferencePaper</dc:type>
  <dc:type>publication-conferencepaper</dc:type>
</oai_dc:dc>
41
11
görüntülenme
indirilme
Görüntülenme 41
İndirme 11
Veri hacmi 2.3 kB
Tekil görüntülenme 40
Tekil indirme 11

Alıntı yap