Konferans bildirisi Açık Erişim
Alshamrani, O. A.; Alrobaish, A. M.; Alqahtani, M. M.; Altakrouni, M. S.; Turhan, E.; Gulmez, G.
An inductance calibration system for standard inductors based on a Maxwell-Wien bridge has been constructed at Saudi Standards Metrology and Quality Organization, National Measurement and Calibration Center (SASO NMCC). Expanded measurement uncertainty has been estimated to be less than 20 mu H/H for 100 mH inductance at 1 kHz. A bilateral comparison between SASO NMCC and TUBITAK UME has been performed in order to evaluate the performance of the calibration system.
| Dosya adı | Boyutu | |
|---|---|---|
|
bib-d407ecda-6c9e-408c-89b4-c1ae26f2ac58.txt
md5:13f6c4d5493c66ede79dc72e86c69e7f |
213 Bytes | İndir |
| Görüntülenme | 41 |
| İndirme | 11 |
| Veri hacmi | 2.3 kB |
| Tekil görüntülenme | 40 |
| Tekil indirme | 11 |