<?xml version='1.0' encoding='utf-8'?> <oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd"> <dc:creator>Celep, Murat</dc:creator> <dc:creator>Danaci, Erkan</dc:creator> <dc:creator>Sakarya, Handan</dc:creator> <dc:creator>Tuncel, Ozlen</dc:creator> <dc:date>2018-01-01</dc:date> <dc:description>TUBITAK UME has established a new power sensor calibration setup incorporating a monitor power sensor utilizing the direct comparison transfer method for frequencies up to 50 GHz. The system adds to TUBITAK UME's existing capability to measure power sensors with type N and 3.5 mm connectors. The new measurement setup was validated using certificate values of power sensors for each working frequency range. Good agreement was found between previous calibration results and those obtained with the new system.</dc:description> <dc:identifier>https://aperta.ulakbim.gov.trrecord/93057</dc:identifier> <dc:identifier>oai:zenodo.org:93057</dc:identifier> <dc:rights>info:eu-repo/semantics/openAccess</dc:rights> <dc:rights>http://www.opendefinition.org/licenses/cc-by</dc:rights> <dc:title>Calibration Factor Measurement up to 50 GHz at TUBITAK UME</dc:title> <dc:type>info:eu-repo/semantics/conferencePaper</dc:type> <dc:type>publication-conferencepaper</dc:type> </oai_dc:dc>
Görüntülenme | 28 |
İndirme | 7 |
Veri hacmi | 1.3 kB |
Tekil görüntülenme | 27 |
Tekil indirme | 7 |