Konferans bildirisi Açık Erişim

Calibration Factor Measurement up to 50 GHz at TUBITAK UME

   Celep, Murat; Danaci, Erkan; Sakarya, Handan; Tuncel, Ozlen

TUBITAK UME has established a new power sensor calibration setup incorporating a monitor power sensor utilizing the direct comparison transfer method for frequencies up to 50 GHz. The system adds to TUBITAK UME's existing capability to measure power sensors with type N and 3.5 mm connectors. The new measurement setup was validated using certificate values of power sensors for each working frequency range. Good agreement was found between previous calibration results and those obtained with the new system.

Dosyalar (185 Bytes)
Dosya adı Boyutu
bib-b625dd20-e180-4a98-a1c7-ae45f8c9f346.txt
md5:6b890268332ea0b3191263fa10399dd5
185 Bytes İndir
28
7
görüntülenme
indirilme
Görüntülenme 28
İndirme 7
Veri hacmi 1.3 kB
Tekil görüntülenme 27
Tekil indirme 7

Alıntı yap