Dergi makalesi Açık Erişim

Realization of photometric base unit of candela traceable to cryogenic radiometer at UME

Samedov, F; Bazkir, O


JSON

{
  "conceptrecid": "97038", 
  "created": "2021-03-16T13:38:11.903208+00:00", 
  "doi": "10.1051/epjap:2005024", 
  "files": [
    {
      "bucket": "5715b0e7-2e91-44da-8074-8a178f222a69", 
      "checksum": "md5:8420342e2ee2f5d5c4ccc2e8f4640bd3", 
      "key": "bib-a6306dc7-a915-4fe8-810f-2933c61ab66e.txt", 
      "links": {
        "self": "https://aperta.ulakbim.gov.tr/api/files/5715b0e7-2e91-44da-8074-8a178f222a69/bib-a6306dc7-a915-4fe8-810f-2933c61ab66e.txt"
      }, 
      "size": 181, 
      "type": "txt"
    }
  ], 
  "id": 97039, 
  "links": {
    "badge": "https://aperta.ulakbim.gov.tr/badge/doi/10.1051/epjap%3A2005024.svg", 
    "bucket": "https://aperta.ulakbim.gov.tr/api/files/5715b0e7-2e91-44da-8074-8a178f222a69", 
    "doi": "https://doi.org/10.1051/epjap:2005024", 
    "html": "https://aperta.ulakbim.gov.tr/record/97039", 
    "latest": "https://aperta.ulakbim.gov.tr/api/records/97039", 
    "latest_html": "https://aperta.ulakbim.gov.tr/record/97039"
  }, 
  "metadata": {
    "access_right": "open", 
    "access_right_category": "success", 
    "communities": [
      {
        "id": "tubitak-adresli-yayinlar"
      }
    ], 
    "creators": [
      {
        "name": "Samedov, F"
      }, 
      {
        "name": "Bazkir, O"
      }
    ], 
    "description": "At National Metrology Institute of Turkey (UME, Ulusal Metroloji Enstitusu), luminous intensity unit of candela was realized using detector-based approach and photometric scale was re-established depending on this new realization. Candela was measured on photometric bench using interferometric distance measurement system and filter-radiometer traceable to UME primary level electrical-substitution cryogenic radiometer. Thermally stabilized filter radiometer, which has been designed for spectral irradiance measurements, is consists of trap detector, filter housing and precision aperture. Different measurement techniques were used to fully characterize each parameter of filter-radiometer; like effective aperture area, spectral transmittance of V (lambda) filter and responsivity of trap detector.", 
    "doi": "10.1051/epjap:2005024", 
    "has_grant": false, 
    "journal": {
      "issue": "3", 
      "pages": "205-213", 
      "title": "EUROPEAN PHYSICAL JOURNAL-APPLIED PHYSICS", 
      "volume": "30"
    }, 
    "license": {
      "id": "cc-by"
    }, 
    "publication_date": "2005-01-01", 
    "relations": {
      "version": [
        {
          "count": 1, 
          "index": 0, 
          "is_last": true, 
          "last_child": {
            "pid_type": "recid", 
            "pid_value": "97039"
          }, 
          "parent": {
            "pid_type": "recid", 
            "pid_value": "97038"
          }
        }
      ]
    }, 
    "resource_type": {
      "subtype": "article", 
      "title": "Dergi makalesi", 
      "type": "publication"
    }, 
    "title": "Realization of photometric base unit of candela traceable to cryogenic radiometer at UME"
  }, 
  "owners": [
    1
  ], 
  "revision": 1, 
  "stats": {
    "downloads": 4.0, 
    "unique_downloads": 4.0, 
    "unique_views": 11.0, 
    "version_downloads": 4.0, 
    "version_unique_downloads": 4.0, 
    "version_unique_views": 11.0, 
    "version_views": 12.0, 
    "version_volume": 724.0, 
    "views": 12.0, 
    "volume": 724.0
  }, 
  "updated": "2021-03-16T13:38:11.949269+00:00"
}
12
4
görüntülenme
indirilme
Görüntülenme 12
İndirme 4
Veri hacmi 724 Bytes
Tekil görüntülenme 11
Tekil indirme 4

Alıntı yap