A Ti/Pt/Co Multilayer Stack for Transfer Function Based Magnetic Force Microscopy Calibrations

Sakar, Baha; Sievers, Sibylle; Scarioni, Alexander Fernandez; Garcia-Sanchez, Felipe; Oztoprak, Ilker; Schumacher, Hans Werner; Ozturk, Osman


JSON-LD (schema.org)

{
  "@context": "https://schema.org/", 
  "@id": 237726, 
  "@type": "ScholarlyArticle", 
  "creator": [
    {
      "@type": "Person", 
      "name": "Sakar, Baha"
    }, 
    {
      "@type": "Person", 
      "affiliation": "Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany", 
      "name": "Sievers, Sibylle"
    }, 
    {
      "@type": "Person", 
      "affiliation": "Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany", 
      "name": "Scarioni, Alexander Fernandez"
    }, 
    {
      "@type": "Person", 
      "affiliation": "Univ Salamanca, Dept Fis Aplicada, Salamanca 37008, Spain", 
      "name": "Garcia-Sanchez, Felipe"
    }, 
    {
      "@type": "Person", 
      "affiliation": "Gebze Tech Univ, Dept Phys, TR-41400 Kocaeli, Turkey", 
      "name": "Oztoprak, Ilker"
    }, 
    {
      "@type": "Person", 
      "affiliation": "Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany", 
      "name": "Schumacher, Hans Werner"
    }, 
    {
      "@type": "Person", 
      "affiliation": "Gebze Tech Univ, Dept Phys, TR-41400 Kocaeli, Turkey", 
      "name": "Ozturk, Osman"
    }
  ], 
  "datePublished": "2021-01-01", 
  "description": "Magnetic force microscopy (MFM) is a widespread technique for imaging magnetic structures with a resolution of some 10 nanometers. MFM can be calibrated to obtain quantitative (qMFM) spatially resolved magnetization data in units of A/m by determining the calibrated point spread function of the instrument, its instrument calibration function (ICF), from a measurement of a well-known reference sample. Beyond quantifying the MFM data, a deconvolution of the MFM image data with the ICF also corrects the smearing caused by the finite width of the MFM tip stray field distribution. However, the quality of the calibration depends critically on the calculability of the magnetization distribution of the reference sample. Here, we discuss a Ti/Pt/Co multilayer stack that shows a stripe domain pattern as a suitable reference material. A precise control of the fabrication process, combined with a characterization of the sample micromagnetic parameters, allows reliable calculation of the sample's magnetic stray field, proven by a very good agreement between micromagnetic simulations and qMFM measurements. A calibrated qMFM measurement using the Ti/Pt/Co stack as a reference sample is shown and validated, and the application area for quantitative MFM measurements calibrated with the Ti/Pt/Co stack is discussed.", 
  "headline": "A Ti/Pt/Co Multilayer Stack for Transfer Function Based Magnetic Force Microscopy Calibrations", 
  "identifier": 237726, 
  "image": "https://aperta.ulakbim.gov.tr/static/img/logo/aperta_logo_with_icon.svg", 
  "license": "http://www.opendefinition.org/licenses/cc-by", 
  "name": "A Ti/Pt/Co Multilayer Stack for Transfer Function Based Magnetic Force Microscopy Calibrations", 
  "url": "https://aperta.ulakbim.gov.tr/record/237726"
}
23
7
görüntülenme
indirilme
Görüntülenme 23
İndirme 7
Veri hacmi 1.6 kB
Tekil görüntülenme 22
Tekil indirme 7

Alıntı yap