Konferans bildirisi Açık Erişim

Radiometric Calibration and SNR Calculation of a SWIR Imaging Telescope

   Yilmaz, Ozgur; Turk, Fethi; Selimoglu, Ozgur

Radiometric calibration of an imaging telescope is usually made using a uniform illumination sphere in a laboratory. In this study, we used the open-sky images taken during bright day conditions to calibrate our telescope. We found a dark signal offset value and a linear response coefficient value for each pixel by using three different algorithms. Then we applied these coefficients to the taken images, and considerably lowered the image non-uniformity. Calibration can be repeated during the operation of telescope with an object that has better uniformity than open-sky. Also SNR (Signal to Noise Ratio) of each pixel was calculated from the open-sky images using the temporal mean and standard deviations. It is found that SNR is greater than 80 for all pixels even at low light levels.

Dosyalar (194 Bytes)
Dosya adı Boyutu
bib-b084018d-f18f-458d-838f-0ffa17d6ec50.txt
md5:8513eff496e99e2923bc69ac84c6aa5d
194 Bytes İndir
29
7
görüntülenme
indirilme
Görüntülenme 29
İndirme 7
Veri hacmi 1.4 kB
Tekil görüntülenme 25
Tekil indirme 7

Alıntı yap