Published January 1, 2019 | Version v1
Conference paper Open

Investigation of Backing Material Used in Ultrasonic Transducers In Terms of Acoustical Parameters

  • 1. TUBITAK Ulusal Metrol Enstitusu TUBITAK UME, Med Metrol Lab, Kocaeli, Turkey

Description

In this study, the backing material located in the bodies of ultrasonic transducers was investigated in terms of hardness values in order to find the best material mixing ratios. In addition, acoustic parameters of materials in the different mixing ratios doped with different amounts of tungsten such as density, sound velocity, attenuation coefficient were measured and acoustic impedance values were calculated. As a consequence, it has been detected that as the weight of tungsten in the mixture boost, acoustic harmonization also rises.

Files

bib-1c9c9f7b-0273-486e-a71b-bb667eb395ce.txt

Files (204 Bytes)

Name Size Download all
md5:16f32118107a979ff04e7d3c38587b26
204 Bytes Preview Download