Investigations of Viscous Heating Effect of Thermocouples under HIFU Applications

  • 1. TUBITAK Ulusal Metrol Enstitusu UME, Med Metrol Lab, Kocaeli, Turkey

Description

In this survey, in order to examine the viscous heating effect of thermocouples, temperature rise due to excitation of a HIFU transducer has been measured by using two different types of T-type thermocouples embedded inside the Tissue Mimicking Material (TMM). In order to obtain the temperature values from the thermocouples more effectively, a PC-based multi-channel data acquisition and monitoring interface were developed by using LabVIEW program. Measurements were carried out inside a TMM which has characteristics like human liver tissue. Temperature rise has been characterized for the different input ultrasonic powers, frequencies and the time intervals applied. The temperature distributions obtained from two different types of T-type thermocouple were compared to each other and discrepancies between these two different thermocouple measurements were shown with the study.