Published January 1, 2005
| Version v1
Journal article
Open
The problem of self-calibration of fluorescence signal in microscale sensor systems
Creators
Description
Most recent developments in fluorescent molecular sensor devices are based on ``ON-OFF'' switching, which is an operation with a single measurable parameter, commonly the quenching of total intensity. In the meantime, with this approach self-calibration on the molecular level is not achievable. This calibration is strongly needed in all kinds of microscale applications, including microarrays, microfluidic systems and living cell imaging. Different possibilities are discussed for overcoming this difficulty and an "OR-OR'' switching concept is suggested that involves the two-channel detection as a promising solution. For achieving the desired efficiency specific conditions are needed: it should be a single reporter dye exhibiting rapid reversible excited-state reaction and providing two-band wavelength ratiometric response.
Files
bib-19c4ba2a-8790-42eb-af35-027a9723ad2f.txt
Files
(139 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:e85eac1010f1933c28f3afd38b135969
|
139 Bytes | Preview Download |