Published January 1, 2005
| Version v1
Journal article
Open
Establishment of a computer-controlled retroreflection measurement system at the National Metrology Institute of Turkey (UME)
Creators
Description
In order to characterize photometric properties of retroreflectors, a fully automated retroreflector measurement system is designed in National Metrology Institute of Turkey (UME). The system is composed of a lighting projector, a goniometer, filter radiometers, 100 dB transimpedance amplifiers, and 24-bit resolution analog-digital converter card with a special software. The established system provides a new calibration capability to determine the luminous intensity and retroreflection coefficients of the retroreflective materials with the expanded uncertainties of 1.07% and 1.13% (k=2), respectively. The traceability in retroreflection measurements was linked to the detector-based photometric scale of UME. (c) 2005 American Institute of Physics.