Simple high-performance liquid chromatographic method for determination of atropine and obidoxime in a parenteral injection device
Oluşturanlar
Açıklama
Atropine and obidoxime in a parenteral injection device are determined by simple HPLC method simultaneously without any pretreatment at 228 nm. The relative standard deviations (R.S.D.) were below 1.6% for the compounds. The correlation coefficient was greater than 0.999 for both compounds in the calibration range. The recoveries at 5 mg/L concentration averaged as 95% for atropine and 102% for obidoxime. The uncertainty of the measurements for atropine and obidoxime was 2.8% and 2.4%, respectively. (C) 2004 Elsevier B.V. All rights reserved.