Published January 1, 2018
| Version v1
Journal article
Open
Rapid and reliable 25-OH vitamin D2 and 25-OH vitamin D3 measurements by multitasker LC-MS/MS
- 1. TUBITAK UME Natl Metrol Inst, Organ Chem Labs, TR-41470 Gebze, Turkey
- 2. Zivak Technol, Gursel Mh Eski Besiktas Cad 44-1-2, TR-34400 Istanbul, Turkey
Description
The 25-OH Vitamin D2 and 25-OH Vitamin D3 measurement method was validated using a Zivak Multitasker Fully Automated Sample Preparation and Injection System coupled to a MS detector. The current method incorporates all sample preparations, SPE clean up procedures, LC and MS facilities into an all in one system. This system can report patient data within 6 minutes including the SPE and Sample preparation and mass spectrometry analysis time. The correlation coefficient of the developed method was greater than 0.99 for measurands, 25-OH-vitamin Ds, in the calibration range. The measurement uncertainty of the measurements were evaluated and found as 7.7 % and 7.6 % for 25-OH vitamin D2 and 25-OH-vitamin D3, respectively.
Files
10-25135-jcm-13-18-02-070.pdf
Files
(767.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:114c5e38af75c0a1578b1c54d5ecc825
|
767.6 kB | Preview Download |