Published January 1, 2015
| Version v1
Journal article
Open
Determination of bromine and tin compounds in plastics using laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS)
Creators
- 1. Uludag Univ, Sci & Art Fac, Dept Chem, TR-16059 Bursa, Turkey
- 2. Sci & Technol Res Council Turkey, Bursa Test & Anal Lab, TR-16190 Bursa, Turkey
Description
The polybrominated flame retardants and organotin compounds were screened in terms of bromine and tin content using laser ablation inductively coupled plasma mass spectrometry in plastics. The calibration standards were prepared using the fused-disk technique, and all samples were investigated under optimal conditions. Using a central composite experimental design, laser parameters, laser energy, pulse rate, scan rate and spot size were identified. The detection limits of the method were 1000 mg kg(-1) and 1600 mg kg(-1) for bromide and tin, whereas the relative standard deviation (%) values of the analysis were 9% and 6% (n=3) for ERM EC681k with 770 +/- 70 mg kg(-1) Br and 86 +/- 6 mg kg(-1) Sn respectively, and 106-115% of Br and 102-104% of Sn were observed for the tetrabromobisphenol A and butyltin trichloride spike plastics, respectively. (C) 2015 Elsevier B.V. All rights reserved.
Files
bib-f2c80e05-84e3-4e7d-9ac5-0799b443dc9f.txt
Files
(191 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:711d67dafaf71c0168893b25379630c2
|
191 Bytes | Preview Download |