Published January 1, 2012
| Version v1
Journal article
Open
Energy resolution improvement of NaI(Tl) scintillation detectors by means of a timing discrimination method
Creators
- 1. Ege Univ, Dept Phys, Fac Sci, TR-35100 Izmir, Turkey
- 2. Ege Univ, Inst Nucl Sci, TR-35100 Izmir, Turkey
Description
Gamma-ray energy spectrum of a Cs-137 radioisotope was obtained by using a NaI(Tl) scintillation detector. Improvement of the energy resolution of the detector was aimed. To achieve this, constant fraction timing method was used. Energy resolution of the photopeak of the isotope was improved from 7.1 to 6.7% through the timing application.