Published January 1, 2015
| Version v1
Journal article
Open
Design and evaluation of a reflectance diffuse optical tomography system
- 1. Akdeniz Univ, Sch Med, Dept Biophys, Biomed Opt Res Unit, TR-07058 Antalya, Turkey
Description
We have designed a continuous wave, back-reflection diffuse optical tomography system; and developed a new practical calibration method including both optode efficiency and positional error corrections. System design, data acquisition and calibration protocols are described in detail. Monte Carlo (MC) simulations of photon distribution for tissue phantoms have been used to obtain the weight matrix to be used in the Rytov approximation to the photon diffusion equation. The system has been evaluated by acquiring data from a tissue phantom with a background scattering coefficient () of and absorption coefficient ( of . An inclusion made of 1 % Intralipid and indocyanine green with and was placed at a 2 cm depth from the tissue phantom surface. After calibration, the average value of the measurements over source-detector pairs at the same distance for each neighborhood was calculated. Perturbation data were obtained by subtracting the average data from the measurements with the same source-detector separation. In the reconstruction, weight matrixes obtained from MC simulation for to were used. The Depth Compensation Algorithm was used in the Tikhonov regularization to identify the location of the inclusion correctly in the reconstruction.
Files
bib-18fa547a-9627-49eb-b850-f51641eee26b.txt
Files
(169 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:368e8a905f576b3f9da98f5dd6ce4f05
|
169 Bytes | Preview Download |