Published January 1, 2020
| Version v1
Journal article
Open
A micropillar-based microfluidic viscometer for Newtonian and non-Newtonian fluids
Creators
- 1. Koc Univ, Grad Sch Biomed Sci & Engn, Istanbul, Turkey
- 2. Duquesne Univ, Dept Engn, Pittsburgh, PA 15219 USA
- 3. Istanbul Tech Univ, Dept Elect & Commun Engn, Istanbul, Turkey
- 4. Koc Univ, Sch Med, Res Ctr Translat Med, Istanbul, Turkey
Description
In this study, a novel viscosity measurement technique based on measuring the deflection of flexible (poly) dimethylsiloxane (PDMS) micropillars is presented. The experimental results show a nonlinear relationship between fluid viscosity and the deflection of micropillars due to viscoelastic properties of PDMS. A calibration curve, demonstrating this nonlinear relationship, is generated, and used to determine the viscosity of an unknown fluid. Using our method, viscosity measurements for Newtonian fluids (glycerol/water solutions) can be performed within 2-100 cP at shear rates gamma = 60.5-398.4 s(-1). We also measured viscosity of human whole blood samples (non-Newtonian fluid) yielding 2.7-5.1 cP at shear rates gamma = 120-345.1 s(-1), which compares well with measurements using conventional rotational vis-cometers (3.6-5.7 cP). With a sensitivity better than 0.5 cP, this method has the potential to be used as a portable microfluidic viscometer for real-time rheological studies. (C) 2020 Elsevier B.V. All rights reserved.
Files
bib-bd76d245-ac92-4dc8-9854-f2cb9048aa1c.txt
Files
(216 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:f7a9287c4fb0953d11104a1ce0fb2b4f
|