A method to enhance coincidence time resolution with applications for medical imaging systems (TOF/PET)
Oluşturanlar
- 1. Ege Univ, Fac Sci, Dept Phys, TR-35100 Izmir, Turkey
- 2. Uludag Univ, Fac Arts & Sci, Dept Phys, Gorukle, Bursa, Turkey
Açıklama
A coincidence timing spectrometer was assembled using NaI(Tl) inorganic and BC418 type organic (plastic) scintillation detectors. The constant fraction timing method was used. Coincidence time resolution values of the detectors, which have great importance in TOF/PET measurements, were obtained separately. The resolutions were enhanced using a different method in a start-stop coincidence spectrometer: the signals from the start detectors were delayed. The results from the FLUKA Monte Carlo simulation code agreed well with the experimental results. (C) 2014 Elsevier Ltd. All rights reserved.