Published January 1, 2016 | Version v1
Journal article Open

Quantitative Analysis of a Promising Cancer Biomarker, Calretinin, by a Biosensing System Based on Simple and Effective Immobilization Process

  • 1. Kirklareli Univ, Sch Hlth, TR-39000 Kirklareli, Turkey
  • 2. Trakya Univ, Fac Sci, Dept Chem, TR-22030 Edirne, Turkey
  • 3. Namik Kemal Univ, Fac Sci, Dept Chem, Tekirdag, Turkey

Description

Calretinin (CAL) is calcium binding protein, and its levels in blood and cerebrospinal fluids are increased, since its expression is increased various cancer types. A novel biosensor system fabricated by immobilization of a specific antibody to CAL, anti-Calretinin (anti-CAL), onto a gold electrode surface via an effective covalent binding method using mercaptohexanol, epichlorohydrin, and ethanolamine was reported for the sensitive, selective, and accurate analysis of CAL. The proposed biosensor showed a linear calibration range between 1 ng/mL and 5 ng/mL. LOD and LOQ values were determined as 0.11 ng/mL and 0.38 ng/mL, respectively. The standard deviation related to the reproducibility of the new biosensor system was calculated as 3.95%. Lastly, in order to state the applicability of the biosensor to early diagnosis of CAL in practice, artificial serum samples spiked with CAL have been analyzed by the proposed biosensor.

Files

bib-da6acdb4-1c9b-459a-b93c-86cc2ed524b4.txt

Files (225 Bytes)

Name Size Download all
md5:290c5d5b78747a23e953290266cb39a8