Published January 1, 2016 | Version v1
Journal article Open

Magnetic micro/nanoparticle flocculation-based signal amplification for biosensing

  • 1. Abdullah Gul Univ, Dept Elect & Elect Engn, BioMINDS Bio Micro Nano Devices & Sensors Lab, Kayseri, Turkey

Description

We report a time and cost efficient signal amplification method for biosensors employing magnetic particles. In this method, magnetic particles in an applied external magnetic field form magnetic dipoles, interact with each other, and accumulate along the magnetic field lines. This magnetic interaction does not need any biomolecular coating for binding and can be controlled with the strength of the applied magnetic field. The accumulation can be used to amplify the corresponding pixel area that is obtained from an image of a single magnetic particle. An application of the method to the Escherichia coli 0157: H7 bacteria samples is demonstrated in order to show the potential of the approach. A minimum of threefold to a maximum of 60-fold amplification is reached from a single bacteria cell under a magnetic field of 20 mT.