Published January 1, 2016 | Version v1
Journal article Open

Computer-Assisted Optimization of Electrodeposited Hydroxyapatite Coating Parameters on Medical Alloys

  • 1. Mustafa Kemal Univ, Fac Arts & Sci, Dept Phys, TR-31040 Antakya, Hatay, Turkey
  • 2. Mustafa Kemal Univ, Fac Arts & Sci, Dept Chem, TR-31034 Antakya, Hatay, Turkey
  • 3. Univ N Texas, Dept Chem, 1155 Union Circle 305070, Denton, TX 76203 USA

Description

CoCrMo bio-metallic alloys were coated with a hydroxyapatite (HA) film by electrodeposition using various electrochemical parameters. Response surface methodology and central composite design were used to optimize deposition parameters such as electrolyte pH, deposition potential, and deposition time. The effects of the coating parameters were evaluated within the limits of solution pH (3.66 to 5.34), deposition potential (-1.13 to -1.97 V), and deposition time (6.36 to 73.64 minutes). A 5-level-3-factor experimental plan was used to determine ideal deposition parameters. Optimum conditions for the deposition parameters of the HA coating with high in vitro corrosion performance were determined as electrolyte pH of 5.00, deposition potential of -1.8 V, and deposition time of 20 minutes.

Files

bib-d3cbaa5a-2a64-492c-914d-b18d98013232.txt

Files (264 Bytes)

Name Size Download all
md5:927be953fd3680054e947a673253b897