A portable microfluidic system for rapid measurement of the erythrocyte sedimentation rate
Creators
- 1. Bilkent Univ, Inst Mat Sci & Nanotechnol, Natl Nanotechnol Res Ctr UNAM, TR-06800 Ankara, Turkey
- 2. Yildirim Beyazit Univ, Fac Med, Ankara, Turkey
Description
The erythrocyte sedimentation rate (ESR) is a frequently used 30 min or 60 min clinical test for screening of several inflammatory conditions, infections, trauma, and malignant diseases, as well as non-inflammatory conditions including prostate cancer and stroke. Erythrocyte aggregation (EA) is a physiological process where erythrocytes form face-to-face linear structures, called rouleaux, at stasis or low shear rates. In this work, we proposed a method for ESR measurement from EA. We developed a microfluidic opto-electro-mechanical system, using which we experimentally showed a significant correlation (R-2 = 0.86) between ESR and EA. The microfluidic system was shown to measure ESR from EA using fingerprick blood in 2 min. 40 mu l of whole blood is filled in a disposable polycarbonate cartridge which is illuminated with a near infrared emitting diode. Erythrocytes were disaggregated under the effect of a mechanical shear force using a solenoid pinch valve. Following complete disaggregation, transmitted light through the cartridge was measured using a photodetector for 1.5 min. The intensity level is at its lowest at complete disaggregation and highest at complete aggregation. We calculated ESR from the transmitted signal profile. We also developed another microfluidic cartridge specifically for monitoring the EA process in real-time during ESR measurement. The presented system is suitable for ultrafast, low-cost, and low-sample volume measurement of ESR at the point-of-care.
Files
bib-7b7a06c7-0e9d-44bb-ba09-5c66461cef62.txt
Files
(171 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:2d1585dbd7e65369a2b170a9905f27c1
|
171 Bytes | Preview Download |