Yayınlanmış 1 Ocak 2016
| Sürüm v1
Dergi makalesi
Açık
A low cost X-ray imaging device based on BPW-34 Si-PIN photodiode
Oluşturanlar
- 1. Istanbul Tech Univ, Fac Sci & Letters, Dept Engn Phys, TR-34469 Istanbul, Turkey
Açıklama
A low cost X-ray imaging device based on BPW-34 silicon PIN photodiode was designed and produced. X-rays were produced from a CEI OX/70-P dental tube using a custom made +/- 30 kV power supply. A charge sensitive preamplifier and a shaping amplifier were built for the amplification of small signals produced by photons in the depletion layer of Si-PIN photodiode. A two dimensional position control unit was used for moving the detector in small steps to measure the intensity of X-rays absorbed in the object to be imaged. An Aessent AES220B FPGA module was used for transferring the image data to a computer via USB. Images of various samples were obtained with acceptable image quality despite of the low cost of the device. (C) 2016 Elsevier B.V. All rights reserved.
Dosyalar
bib-4def9e1b-d4b8-408d-a0f7-6d22112b3507.txt
Dosyalar
(271 Bytes)
| Ad | Boyut | Hepisini indir |
|---|---|---|
|
md5:6b49b48327a583eefad8378fbb333945
|
271 Bytes | Ön İzleme İndir |