Considerations in applying compressed sensing to in vivo phosphorus MR spectroscopic imaging of human brain at 3T

  • 1. Bogazici Univ, Biomed Engn Inst, Rasathane Cad,Kandilli Campus,Kandilli Mah, TR-34684 Istanbul, Turkey

Description

The purpose of this study was to apply compressed sensing method for accelerated phosphorus MR spectroscopic imaging (P-31-MRSI) of human brain in vivo at 3T. Fast P-31-MRSI data of five volunteers were acquired on a 3T clinical MR scanner using pulse-acquire sequence with a pseudorandom undersampling pattern for a data reduction factor of 5.33 and were reconstructed using compressed sensing. Additionally, simulated P-31-MRSI human brain tumor datasets were created to analyze the effects of k-space sampling pattern, data matrix size, regularization parameters of the reconstruction, and noise on the compressed sensing accelerated P-31-MRSI data. The P-31 metabolite peak ratios of the full and compressed sensing accelerated datasets of healthy volunteers in vivo were similar according to the results of a Bland-Altman test. The estimated effective spatial resolution increased with reduction factor and sampling more at the k-space center. A lower regularization parameter for both total variation and L1-norm penalties resulted in a better compressed sensing reconstruction of P-31-MRSI. Although the root-mean-square error increased with noise levels, the compressed sensing reconstruction was robust for up to a reduction factor of 10 for the simulated data that had sharply defined tumor borders. As a result, compressed sensing was successfully applied to accelerate P-31-MRSI of human brain in vivo at 3T.

Files

bib-46f7d9c5-2d90-49ee-8bee-532218788ab7.txt

Files (228 Bytes)

Name Size Download all
md5:637960827130c6341d32775fb992ccc3
228 Bytes Preview Download