Published January 1, 2017 | Version v1
Journal article Open

Analysis of reconstruction performance of magnetic resonance conductivity tensor imaging (MRCTI) using simulated measurements

  • 1. Mersin Univ, Dept Elect & Elect Engn, Fac Engn, Mersin, Turkey
  • 2. Middle East Tech Univ, Fac Engn, Dept Elect & Elect Engn, Ankara, Turkey

Description

Magnetic resonance conductivity tensor imaging (MRCTI) was proposed recently to produce electrical conductivity images of anisotropic tissues. Similar to magnetic resonance electrical impedance tomography (MREIT), MRCTI uses magnetic field and boundary potential measurements obtained utilizing magnetic resonance imaging techniques. MRCTI reconstructs tensor images of anisotropic conductivity whereas MREIT reconstructs isotropic conductivity images. In this study, spatial resolution and linearity of five recently proposed MRCTI algorithms are evaluated using simulated measurements gathered from three different computer models. The results show that all five algorithms have quite similar reconstruction performances. Since the ABz S algorithm is easier to apply compared to the other four algorithms it can be said to be the best algorithm among the five algorithms.

Files

10-3906-elk-1409-97.pdf

Files (378.1 kB)

Name Size Download all
md5:a9f5740bddcb3d20e709ec9b963b7f8a