Published January 1, 2008
| Version v1
Journal article
Open
A Magnetically Actuated Resonant Mass Sensor With Integrated Optical Readout
Creators
- 1. Koc Univ, Coll Engn, TR-34450 Istanbul, Turkey
Description
Nickel cantilevers with integrated diffraction gratings are used as resonant mass sensors with a resolution of 500 femtograms. Their applicability to biosensing is demonstrated with human opioid receptors. The device is fabricated through a single-mask lithographic process. The microoptical readout provides a simple measurement platform with one external photodiode. Thanks to its ac operation principle, the device is immune to environmental noise and entails a high tolerance to fabrication defects. Obtained signal-to-noise ratio is comparable to that of a high-end Doppler vibrometer. The device with these aspects for systems integration and microarray technology is a candidate for low-cost portable sensors.