Published January 1, 2018
| Version v1
Conference paper
Open
A Read-Out System for Disposable Paper-Based Sensors
Creators
- 1. Istanbul Bilgi Univ, Dept Elect & Elect Engn, Istanbul, Turkey
Description
This study presents design, implementation and test of a read-out system for disposable paper-based sensors. Proposed system basically consists of a Strathmore (R) 300 series Bristol paper-based cantilever with the dimension of lem x 3.5cra and the thickness of roughly 400/art which is covered with a thin piezoresistive layer and a read-out circuitry that is composed of a Wheatstone Bridge circuit and instrumentation amplifier (INA128). Proposed read-out system is simulated, implemented and tested. The sensitivity of the proposed read-out system is measured as 0.25m/V/mu m and Delta V-out and 1.5m Omega/mu m for Delta R-PR.
Files
bib-f57a0b82-9584-4e1c-91a6-d53a590b015b.txt
Files
(161 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:a3931ef9bf44e6f7f5a21d0fc6f68f58
|
161 Bytes | Preview Download |