Published January 1, 2026 | Version v1
Journal article Open

Tunable Acoustic Layers Enable Gel-Free, Conformal Wearable Ultrasound

  • 1. Koc Univ, Dept Biomed Sci & Engn & Mech Engn, Istanbul, Turkiye
  • 2. Koc Univ, Dept Mech Engn & Biomed Sci & Engn & Chem, Istanbul, Turkiye
  • 3. Koc Univ, Dept Chem, Istanbul, Turkiye

Description

Rigid probes and coupling gels still tether diagnostic ultrasound to the clinic, precluding continuous use on moving, curved, sweating human skin. We report a tunable design platform based on epoxy-polyetheramine composites enabling independently tuned matching and backing layers, spanning ranges of acoustic impedance (2.0-5.4 MRayl), attenuation (12-117 dBcm-1 at 2.25 MHz), and Young's modulus (4-4700 MPa). A gradient-impedance matching layer lifts fractional bandwidth from 15% to 54%, while the visco-elastic backing boosts signal-to-noise-ratio (SNR) by 9 dB without sacrificing conformity. In vivo validation through blood pressure measurements demonstrates the efficacy of these tailor-designed transducers in capturing key hemodynamic parameters, enabling continuous cardiovascular monitoring beyond traditional clinical settings. This versatile platform translates polymer network chemistry into a design process, offering a general path to gel-free wearable ultrasound and, more broadly, to on-skin acoustic, photo-acoustic, and therapeutic devices.

Files

bib-385925cc-dcdf-4e6a-8749-5a32b01d347e.txt

Files (198 Bytes)

Name Size Download all
md5:6a2c725bace27cd1231ec6fefe154216