Graphene-Based Nanomaterial Electrochemical Biosensors for the Determination of the H3N2 Virus
Creators
- 1. Mugla Sitki Kocman Univ, Res Lab Ctr, Kotekli Mugla, Turkiye
Description
Since influenza and COVID-19 cause similar symptoms, the need for selective, sensitive and practical systems for influenza A detection has been highlighted. For this reason, in this work, a graphene (Gr) based nanomaterial modified anti-hemagglutinine 3, from now on will be called Gr-antiH3, based disposible electrochemical influenza immunosensor was developed. In order to obtain the highest sensitivity, the effects of various Gr-based nanomaterials on developed electrochemical immunosensor responses were investigated. For this purpose, graphene oxide, Gr, Gr-platinum (Gr-Pt), Gr-gold and Gr-gold-platinum nanomaterials were synthesized and characterized via electrochemical impedance spectroscopy (EIS). As a result, since the best results were obtained with Gr-Pt, this nanomaterial was integrated into H3N2 immunosensor system as a modifier. The Gr-Pt modified immunosensor was shown to provide a linear range from 0.5 mu g/mL to 7 mu g/mL. The limit of detection was 0.38 mu g/mL and limit of quantification was 1.28 mu g/mL with a relative standard deviation of 3.05%. The developed biosensor was applied to H3N2 detection in synthetic saliva samples and promising results were obtained with recoveries of 100.41 +/- 1.10%.
Files
bib-01e5c15a-c717-48ad-9d8f-4a7ea9fb094e.txt
Files
(159 Bytes)
| Name | Size | Download all |
|---|---|---|
|
md5:a587a0910824a6409309d461793e7bf7
|
159 Bytes | Preview Download |