Published January 1, 2023 | Version v1
Journal article Open

Investigation of Alpha Detection Performances of Two Different Types of Photomultiplier Tubes

  • 1. Izmir Univ Econ, Fac Arts & Sci, Dept Phys, Izmir, Turkiye
  • 2. Ege Univ, Fac Sci, Dept Phys, Izmir, Turkiye

Description

Alpha particle detection performances of a Hamamatsu R1828-01 model and a RCA 6342A model photomultiplier tubes (PMT) were tested. For the sensitivity of the PMTs to alpha particles, an Eljen AJ-440 model ZnS(Ag) scintillator material in sheet form was used in front of their windows. To test their performances, the alpha peaks of an Am-241 source and the energy resolution values of the detectors that composed of these PMTs were compared. It has been showed that R1828-01 model PMT utilized for fast timing measurements in radiation detection could be used in alpha particle detection efficiently. This work revealed its another advantageous feature in addition to existing specifications.