Yayınlanmış 1 Ocak 2021 | Sürüm v1
Dergi makalesi Açık

High-sensitivity biomedical sensor based on photoacoustic and cavity enhanced absorption spectroscopy with a new software platform for breath analysis

  • 1. Aksaray Univ, Elect & Elect Engn, Aksaray, Turkey
  • 2. Suleyman Demirel Univ, Biomed Engn, Isparta, Turkey

Açıklama

An easy-to-use highly sensitive sensor is reported for trace gas analysis. A near-infrared fiber-coupled external cavity diode laser in combination with a photoacoustic spectroscopy cell and a cavity enhanced absorption spectroscopy cell is used for analysis of trace gases. A software platform for direct absorption and wavelength modulation spectroscopy is developed in order to identify and quantify the molecules and to achieve a higher signal-to-noise ratio. Considering all of these features, the main advantage of our system is to analyze two different samples simultaneously and quickly. Trace gas measurement is assessed, and a detection limit of 1.5 ppb at 6528.76 cm(-1) for ammonia is demonstrated. Furthermore, the sensor with our software platform can be easily used outside of the laboratory, for example, in hospitals. (C) 2021 Optical Society of America

Dosyalar

bib-b93254fe-244d-4712-82f8-a4bd4f3304de.txt

Dosyalar (228 Bytes)

Ad Boyut Hepisini indir
md5:001615eab388705f926223ef85946266
228 Bytes Ön İzleme İndir