High-sensitivity biomedical sensor based on photoacoustic and cavity enhanced absorption spectroscopy with a new software platform for breath analysis
Creators
- 1. Aksaray Univ, Elect & Elect Engn, Aksaray, Turkey
- 2. Suleyman Demirel Univ, Biomed Engn, Isparta, Turkey
Description
An easy-to-use highly sensitive sensor is reported for trace gas analysis. A near-infrared fiber-coupled external cavity diode laser in combination with a photoacoustic spectroscopy cell and a cavity enhanced absorption spectroscopy cell is used for analysis of trace gases. A software platform for direct absorption and wavelength modulation spectroscopy is developed in order to identify and quantify the molecules and to achieve a higher signal-to-noise ratio. Considering all of these features, the main advantage of our system is to analyze two different samples simultaneously and quickly. Trace gas measurement is assessed, and a detection limit of 1.5 ppb at 6528.76 cm(-1) for ammonia is demonstrated. Furthermore, the sensor with our software platform can be easily used outside of the laboratory, for example, in hospitals. (C) 2021 Optical Society of America