Published January 1, 2011 | Version v1
Journal article Open

Design of a Centrifugal Blood Pump: Heart Turcica Centrifugal

  • 1. Koc Univ, Dept Mech Engn, Mfg & Automat Res Ctr, TR-34450 Istanbul, Turkey
  • 2. Florence Nightingale Hosp, Dept Cardiovasc Surg, Istanbul, Turkey

Description

A prototype of a new implantable centrifugal blood pump system named Heart Turcica Centrifugal (HTC) was developed as a left ventricular assist device (LVAD) for the treatment of end-stage cardiac failure. In the development of HTC, effects of blade height and volute tongue profiles on the hydraulic and hemolytic performances of the pump were investigated. As a result, the prototype was manufactured using the best blade height and volute tongue profiles. Performance of the prototype model was experimentally evaluated in a closed-loop flow system using water as the medium. The hydraulic performance requirement of an LVAD (5 L/min flow rate against a pressure difference of 100 mm Hg) was attained at 2800 rpm rotational speed.

Files

bib-57187774-86cd-492d-a06a-f5e1cf486261.txt

Files (158 Bytes)

Name Size Download all
md5:29e865f29f80a9e1241a1287f7468e86
158 Bytes Preview Download