Yayınlanmış 1 Ocak 2018 | Sürüm v1
Dergi makalesi Açık

Solid-Phase Microextraction and Determination of Tin Species in Beverages and Food Samples by Using Poly (epsilon-Caprolactone-b-4-Vinyl Benzyl-g-Dimethyl Amino Ethyl Methacrylate) Polymer in Syringe System: a Multivariate Study

  • 1. Univ Sindh, Inst Adv Res Studies Chem Sci, Jamshoro, Pakistan

Açıklama

A new solid-phase microextraction (SP mu E) procedure has been developed for separation and preconcentration of Sn ions by using graphite furnace atomic absorption spectrometry (GFAAS). This technique is based on the complexation of Sn(IV) ions with 1-(2-pyridylazo)-2-naphthol (PAN). The poly (epsilon-caprolactone-b-4-vinyl benzyl-g-dimethyl amino ethyl methacrylate) polymer (PCL vacr) was used as an adsorbent, and it was loaded in micropipette tip of syringe system. Sn(IV) ions were adsorbed on polymer at pH 6. Different experimental conditions were optimized such as pH, amount of complexing agent, and amount of adsorbent. The detection limit (LOD), limit of quantification, preconcentration factor (PF), and relative standard deviation (RSD) were found as 4.5 ng L-1, 13.5 ng L-1, 100, and 3.3%, respectively. Certified reference materials were used to confirm the accuracy of the investigated procedure, and the procedure was successfully practiced for determination of entire concentration of tin within beverages and different food samples.

Dosyalar

bib-ad595cca-5283-4378-b457-c445d8876a66.txt

Dosyalar (327 Bytes)

Ad Boyut Hepisini indir
md5:e6b91ca3d0e89d76b79773c31e7d5a29
327 Bytes Ön İzleme İndir