\WERS {Evaluation of Energy Resolution by Changing Angle and Position of Incident Photon in a LYSO Calorimeter}

  • 1. Department of Physics, Bursa Uludag University, 16059 Bursa, Turkey

Description

In this paper, we investigate the effect on energy resolution from changing the angle and the position of incidence photon for a \(5 \times 5\) crystal matrix built with (\(25 \times 25 \times 200\)) mm\(^{3}\) LYSO scintillating crystals. Those crystals have been proposed for the electromagnetic calorimeter of the Turkish Accelerator Center-Particle Factory (TAC-PF) detector. The energy resolution was obtained as \(\sigma _{E}/E = 0.42\% / \sqrt {E/{\mathrm {GeV}}} \oplus 1.60\%\) at the center of the matrix in the energy range of 50 MeV to 2 GeV. When we examined the dependence of the energy resolution on the incidence angle of the photon, resolution began to deteriorate at angles greater than \(2^{\circ }\) on the \(5 \times 5\) crystal matrix. Moreover, energy resolution at the corners of the central crystal was worse than at the center of the central crystal by a factor of 1.3 at 50~MeV and 1.1 at 2 GeV.