Published January 1, 1998 | Version v1
Conference paper Open

Intercomparison of a 10 mH inductance standard between UME-PTB

Description

Turkish and German Metrology Institutes (UME and PTB) performed an intercomparison of 10 mH inductance standards at 1 kHz. The measurements of the thermostatically-controlled transfer standard of PTB were taken by new Maxwell-Wien bridge with a Wagner arm made in UME. Relative difference of UME and PTB measurements on the same standard was found 5.10(-6). Results have shown that this type of Maxwell-Wien bridge is capable of producing very precise measurements on standard inductors and promised further developments that will increase the accuracy of the system.

Files

bib-e9c9ece3-10ad-4da2-8cad-4cc86e2c3b78.txt

Files (197 Bytes)

Name Size Download all
md5:acbf7a03528cde111d99e66710d2c3d6
197 Bytes Preview Download