Zero gas reference standards

  • 1. Natl Phys Lab, Environm Div, Hampton Rd, Teddington TW11 0LW, Middx, England
  • 2. VSL, Thijsseweg 11, NL-2629 JA Delft, Netherlands
  • 3. LNE, 1 Rue Gaston Boissier, F-75015 Paris, France
  • 4. DMDM, Grp Opt & Phys Chem Quant, Reference Mat Lab, Mike Alasa 14, Belgrade 11000, Serbia
  • 5. TUBITAK UME Natl Metrol Inst, Gas Metrol Lab, PK 54, TR-41470 Gebze, Kocaeli, Turkey

Açıklama

High purity nitrogen or air, often referred to as zero gas, is essential for calibrating instrumentation used in a wide variety of measurements, such as air quality monitoring. Providing traceable and accurate quantification of impurities of the target component in zero gas is a major challenge, as the detection limits of appropriate analytical techniques are often close to the amount fraction of the measurand and accurate reference standards to underpin these measurements are yet to be developed. Purity analysis of zero gas is a significant contributor to the uncertainty budget and is compromised by a dependence on a zero reference for its realisation. This work describes state-of-the-art methods currently used to certify zero gas standards. We distinguish between absolute and relative techniques and provide guidance on calibration of the latter at trace amount fractions. Measurements to assess the performance of a selection of zero air generators and purifiers are also presented here.