Dergi makalesi Açık Erişim

Validation and sensitivity evaluation of the ID-GC-TOF-MS method for determination of PAHs in biogas

   Gokcen, Taner; Bilsel, Gokhan; Tarhan, Tanil; Un, Ilker; Goren, Ahmet C.; van der Veen, Adriaan

In this study, an ID-GC-TOF-MS method was developed and validated for polycyclic aromatic hydrocarbons (PAHs) by spiking solutions onto thermal desorption tubes using calibration solution loading ring (CSLR) apparatus and sweeping them by the aid of air stream. Once the method was developed and validated, the effect of other compounds found in biogas and biomethane was investigated by the same methodology used in method development by using a gravimetrically prepared biogas mixture consisting of 45% methane, 40% carbon dioxide and 15% nitrogen. There are standard reference methods for measurements of PAHs in air but there was no available standard method for PAHs in biogas and biomethane yet. This study provides a basis for the future studies to develop a standard method. Data obtained through this study showed that present method is not affected by the compounds found in biogas that are not present in other matrices such as ambient air.

Dosyalar (1.2 MB)
Dosya adı Boyutu
10-25135-jcm-11-17-11-01.pdf
md5:b8e09d5e2c034e212aa1b4c3e9cf1690
1.2 MB İndir
65
33
görüntülenme
indirilme
Görüntülenme 65
İndirme 33
Veri hacmi 39.8 MB
Tekil görüntülenme 58
Tekil indirme 33

Alıntı yap