Evaluation of expanded uncertainties in luminous intensity and illuminance calibrations
Creators
Description
Detector-based calibrating methods and expressions for calculation of photometric uncertainties related to uncertainties in the calibrations of luminous intensity of a light source, illuminance responsivity of a photometer head, and calibration factors of an illuminance meter are discussed. These methods permit luminous intensity calibrations of incandescent light sources, luminous responsivity calibrations of photometer heads, and calibration factors of illuminance meters to be carried out with relative expanded uncertainties (with a level of confidence of 95.45%) of 0.4%, 0.4%, and 0.6%, respectively (C) 2008 Optical Society of America