Yayınlanmış 1 Ocak 2018 | Sürüm v1
Dergi makalesi Açık

Voltammetric Determination of Boron using Poly Xylenol Orange-Modified Pencil Graphite Electrode

  • 1. Natl Metrol Inst TUBITAK UME, Chem Grp, Electrochem Lab, Kocaeli, Turkey
  • 2. Balikesir Univ, Dept Chem, Fac Arts & Sci, Balikesir, Turkey

Açıklama

This study describes a new voltammetric method for sensitive and accurate determination of boron using the poly xylenol orange-modified and oxidized pencil graphite electrode. The oxidation of tiron at +0.86 V in the boron-tiron complex formed in phosphate buffer solution at pH 8 was evaluated as a response. The parameters for voltammetric determination of boron were optimized. The modified electrode was characterized by cyclic voltammetry, electrochemical impedance spectroscopy, scanning electron microscopy, and energy-dispersive X-ray spectroscopy. The limits of detection and quantification were estimated to be 28 and 83 mu g/L, respectively. The relative standard deviation value was 4.89% for 0.2 mg/L boron. The recovery results were 98.72, 100.53, and 100.14% for tap water, drinking water, and eye lotion samples, respectively. The accuracy of the method was tested using a certified reference material (UME CRM 1201: Elements in Spring Water).

Dosyalar

bib-3e8d98b9-3fcf-4605-8efe-42abdbcc31db.txt

Dosyalar (167 Bytes)

Ad Boyut Hepisini indir
md5:aebbcbe16a98ed343984fa50f5655208
167 Bytes Ön İzleme İndir