Comparison of Blackbodies for Calibration of Infrared Ear Thermometers

  • 1. Univ Ljubljana, Metrol Inst Republ Slovenia, Fac Elect Engn, Lab Metrol & Qual MIRS UL FE LMK, Ljubljana 1000, Slovenia
  • 2. Riso DTU, DK-4000 Roskilde, Denmark
  • 3. Lab Commun Metrol LNE Cnam, F-75724 Paris 15, France
  • 4. PTB, D-10587 Berlin, Germany
  • 5. TUBITAK Ulusal Metrol Enstitusu UME, TR-41470 Gebze, Kocaeli, Turkey
  • 6. Natl Phys Lab, Teddington TW11 0LW, Middx, England
  • 7. Justervesenet, N-2007 Kjeller, Norway
  • 8. Swiss Fed Off Metrol METAS, CH-3003 Bern, Switzerland
  • 9. NMi Van Swinden Lab BV, NL-2629 JA Delft, Netherlands

Açıklama

The article presents the results of the EURAMET Project No. 927 "Comparison of blackbodies for calibration of infrared ear thermometers (IRETs)". The objective of the comparison was to determine the agreement of blackbodies used for the calibration of IRETs among European national laboratories. To verify the accuracy of an IRET, a suitable blackbody (BB) is needed. Such a blackbody related to the EN standard, designed for the calibration of ear thermometers and immersed in a stirred water bath, was provided for the comparison by the pilot laboratory. The pilot provided also the transfer IRET and organized the comparison.